A comparison of exhaled nitric oxide measurements performed using three different analysers.

AbstractINTRODUCTION:
Exhaled nitric oxide (NO) is an established technique for monitoring airway inflammation. We have compared exhaled NO measurements from 3 different analysers; Ecomedics (E), Niox (N) and Logan (L).
METHODS:
Thirty subjects (10 non-smoking healthy subjects, 10 non-smoking patients with asthma and 10 ex-smoking COPD patients) performed 3 repeated measurements of exhaled NO at a flow rate of 50 ml/s on each of the 3 analysers. Within analyser variability was determined by calculating the repeatability coefficient for each analyser. Differences between analysers were assessed by (1) the differences between group means and (2) the Bland Altman method to estimate the variability expected for an individual using the 3 analysers.
RESULTS:
The repeatability coefficients (expressed as ratios) were 1.12, 1.19 and 1.19 for N, E and L, respectively. There were significant differences (P<0.05) between analysers; the Logan analyser gave the highest group mean values and Ecomedics gave the lowest group mean values. Differences between analysers were observed in all subject groups (healthy, asthma, COPD). Similar results were obtained in the 3 groups when analysed separately. Bland Altman analysis gave the following ratios [data are mean ratio (95% limits of agreement)]; N:E 1.59 (1.02-2.50), L:N 1.23 (0.72-2.13), L:E 1.96 (1.09-3.57).
CONCLUSION:
Our findings indicate that exhaled NO measurements in healthy subjects and patients with airways disease differ according to the type of analyser used.
AuthorsZoë Borrill, David Clough, Nick Truman, Julie Morris, Stephen Langley, Dave Singh
JournalRespiratory medicine (Respir Med) Vol. 100 Issue 8 Pg. 1392-6 (Aug 2006) ISSN: 0954-6111 [Print] England
PMID16431095 (Publication Type: Comparative Study, Journal Article)
Chemical References
  • Nitric Oxide
Topics
  • Adult
  • Aged
  • Asthma (metabolism)
  • Breath Tests (instrumentation)
  • Exhalation
  • Female
  • Humans
  • Male
  • Nitric Oxide (analysis)
  • Reproducibility of Results

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: