[Simple method to estimate object to image magnification and aniseikonia following cataract surgery].

AbstractBACKGROUND AND PURPOSE:
Aniseikonia as one of the major risk factors for asthenopic problems is mostly overlooked in modern cataract surgery. The purpose of this study was to develop a simple calculation scheme for clinicians to predict the object to image magnification in a pseudophakic eye with biometric data.
METHODS:
The calculation scheme for object to image magnification in the pseudophakic eye is based on a vergence calculation of the lens power with theoretical optical formulae. From the biometric data, which are typically derived from both eyes during lens power calculation, the vergences in front of and behind the 3 or 4 refractive surfaces of the pseudophakic eye model are used to predict the magnification for objects at infinity or objects located at a finite measurement distance (e.g. 5 m).
RESULTS:
With a formula-based lens power calculation a pseudophakic eye model is set up with 3 or 4 refractive surfaces (postoperative spectacle refraction; thick cornea described by anterior surface or thick cornea characterized by anterior and posterior surfaces; intraocular lens). The vergence in front of and behind each refractive surface is derived by means of linear Gaussian optics. The quotient of the product of all vergences in front of the surfaces and the product of all vergences behind the respective surfaces describes the object to image magnification of the eye. A comparison of the object to image magnification of both eyes yields the retinal image size disparity or aniseikonia. This calculation strategy is shown in a step-by-step approach exemplarily for the Haigis and Hoffer‑Q formulae (3 surfaces) and the Castrop formula (4 surfaces).
CONCLUSION:
If during planning and lens power calculation biometry is performed for both eyes, ocular magnification of both eyes can be easily derived with this calculation scheme and aniseikonia can be extracted from a comparison of magnification of both eyes. Such a simple prediction should be established as a standard for precataract biometry and lens power calculation for early detection and avoidance of asthenopic complaints after cataract surgery.
AuthorsAchim Langenbucher, Peter Hoffmann, Jascha Wendelstein, Nóra Szentmáry
JournalDer Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft (Ophthalmologe) Vol. 119 Issue Suppl 1 Pg. 56-63 (Jan 2022) ISSN: 1433-0423 [Electronic] Germany
Vernacular TitleEinfaches Verfahren zur Abschätzung des postoperativen Abbildungsmaßstabs und der Aniseikonie bei der Kataraktoperation.
PMID34061224 (Publication Type: Journal Article)
Copyright© 2021. The Author(s).
Topics
  • Aniseikonia (diagnosis)
  • Biometry
  • Cataract
  • Cataract Extraction
  • Humans
  • Lenses, Intraocular
  • Refraction, Ocular

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: