HOMEPRODUCTSCOMPANYCONTACTFAQResearchDictionaryPharmaSign Up FREE or Login

[MGuard Dacron mesh-covered stent implantation in patients with ST-elevation myocardial infarction and high thrombotic burden: in-hospital and long-term outcome in a single high-volume center].

AbstractBACKGROUND:
The MGuard (MG) is a stainless-steel closed cell stent covered with an ultra-thin polymer mesh sleeve, which has the purpose of ensuring reduced distal embolization of thrombotic material, thus lowering the risk of no-reflow/slow-flow phenomena. Only few data are available that evaluated the usefulness of the MG stent in patients with ST-elevation myocardial infarction (STEMI) with a high thrombotic burden.
METHODS:
We prospectively collected data of patients presenting with STEMI and high thrombotic burden (thrombus burden grade 4 or 5 according to the TIMI score) who underwent primary percutaneous coronary intervention (PCI) with MG stent implantation in our center. Lesions involving a bifurcation or very calcified and tortuous vessels were not included. Final TIMI 3 flow, normal myocardial blush grade (MBG), and complete ST-segment resolution were defined as short-term endpoints, whereas major adverse cardiovascular events were evaluated during follow-up.
RESULTS:
From August 2008 to June 2013, the MG stent was implanted in 104 (9.3%) of 1108 patients who underwent primary PCI. Cardiogenic shock on admission was present in 4 patients (3.8%). Final TIMI 3 flow was achieved in 97.1% of patients, MBG 3 in 56.7%, and regression of ST-segment elevation ≥70% within 90 min was recorded in 64.4% of patients. In-hospital mortality was 2.9% (n=3), and at a mean follow-up of 455±430 days overall major adverse cardiovascular events were 11 (10.6%). Cardiac death occurred in 5 patients (4.9%) and stent thrombosis in 2 (1.9%).
CONCLUSIONS:
In selected patients with STEMI undergoing primary PCI, the use of the mesh-covered MG stent in vessels with high thrombotic burden despite thrombus aspiration was effective and safe both immediately and at mid-term follow-up.
AuthorsEnrico Cerrato, Cristina Rolfo, Francesco Tomassini, Nicolò Montali, Alfonso Gambino, Vincenzo Infantino, Sara Palacio Restrepo, Denise Baricocchi, Rosa Nevola, Andrea Gagnor, Ferdinando Varbella
JournalGiornale italiano di cardiologia (2006) (G Ital Cardiol (Rome)) Vol. 16 Issue 6 Pg. 373-9 (Jun 2015) ISSN: 1827-6806 [Print] Italy
Vernacular TitleMGuard "Dacron mesh covered" stent per il trattamento della sindrome coronarica acuta con sopraslivellamento del tratto ST in presenza di elevato carico trombotico: outcome intraospedaliero e a lungo termine in un singolo centro ad alto volume.
PMID26156698 (Publication Type: Journal Article)
Chemical References
  • Anticoagulants
  • Polyethylene Terephthalates
Topics
  • Aged
  • Anticoagulants (therapeutic use)
  • Combined Modality Therapy
  • Coronary Angiography
  • Coronary Circulation
  • Coronary Thrombosis (complications, diagnostic imaging, drug therapy, surgery)
  • Equipment Design
  • Female
  • Hospitals, Special
  • Humans
  • Male
  • Middle Aged
  • Myocardial Infarction (etiology, physiopathology, surgery)
  • Percutaneous Coronary Intervention (adverse effects, instrumentation)
  • Polyethylene Terephthalates
  • Postoperative Complications (etiology, prevention & control)
  • Prospective Studies
  • Severity of Illness Index
  • Stents (adverse effects)
  • Surface Properties
  • Thromboembolism (etiology, prevention & control)
  • Treatment Outcome