Abstract |
The aim of this study was to investigate whether cyclic fatigue resistance is increased for nickel titanium instruments manufactured with improved heating processes in clockwise or counterclockwise continuous rotation. The instruments compared were produced either using the R-phase heat treatment (K3XF; SybronEndo, Orange, CA, USA) or the M-wire alloy (ProFile Vortex; DENTSPLY Tulsa Dental Specialties, Tulsa, OK, USA). Tests were performed with a specific cyclic fatigue device that evaluated cycles to failure of rotary instruments in curved artificial canals. Results indicated no significant difference in resistance to cyclic fatigue when rotary nickel titanium instruments are used in clockwise or counterclockwise continuous rotation. In both directions of rotation, size 04-25 K3XF showed a significant increase (P < 0.05) in the mean number of cycles to failure when compared with size 04-25 ProFile Vortex.
|
Authors | Gianlucca Gambarini, Richard Gergi, Nicola Maria Grande, Nada Osta, Gianluca Plotino, Luca Testarelli |
Journal | Australian endodontic journal : the journal of the Australian Society of Endodontology Inc
(Aust Endod J)
Vol. 39
Issue 3
Pg. 151-4
(Dec 2013)
ISSN: 1747-4477 [Electronic] Australia |
PMID | 24279663
(Publication Type: Comparative Study, Journal Article)
|
Copyright | © 2012 The Authors. Australian Endodontic Journal © 2012 Australian Society of Endodontology. |
Chemical References |
- Dental Alloys
- titanium nickelide
- Nickel
- Titanium
|
Topics |
- Dental Alloys
(chemistry)
- Dental Pulp Cavity
(anatomy & histology)
- Equipment Design
- Equipment Failure
- Hot Temperature
- Humans
- Materials Testing
- Metallurgy
- Models, Anatomic
- Nickel
(chemistry)
- Root Canal Preparation
(instrumentation)
- Rotation
- Stress, Mechanical
- Titanium
(chemistry)
- Torque
|