Comparison of algorithms to suppress artifacts from the natural lens in fluorescence lifetime imaging ophthalmoscopy (FLIO).

Abstract
Fluorescence lifetime imaging ophthalmoscopy (FLIO) has developed as a new diagnostic tool in ophthalmology. FLIO measurements are taken from 30° retinal fields in two spectral channels (short spectral channel (SSC): 498-560 nm, long spectral channel (LSC): 560-720 nm). Because of the layered structure of the eye, the detected signal is an interaction of the fluorescence decay of the anterior part and of the fundus. By comparing FLIO measurements before and after cataract surgery, the impact of the natural lens was proven, despite the application of a confocal laser scanning (cSLO) technique. The goal of this work was to determine the best algorithmic solution to isolate the sole fundus fluorescence lifetime from the measured signal, suppressing artifacts from the natural lens. Three principles based on a tri-exponential model were investigated: a tailfit, a layer-based approach with a temporally shifted component, and the inclusion of a separately measured fluorescence decay of the natural lens. The mean fluorescence lifetime τm,12 is calculated using only the shortest and the intermediate exponential component. τm,all is calculated using all three exponential components. The results of tri-exponential tailfit after cataract surgery were considered as a reference, because the implanted artificial lens can be assumed as non-fluorescent. In SSC, the best accordance of τm,all of the reference was determined with τm,12 of the tailfit before surgery. If high-quality natural lens measurements are available, the correspondence of τm,12 is best with τm,all of the reference. In LSC, there is a good accordance for all models between τm,12 before and after surgery. To study the pure fundus fluorescence decay in eyes with natural lenses, we advise to utilize fluorescence lifetime τm,12 of a triple-exponential tailfit, as it corresponds well with the mean fluorescence lifetime τm,all of eyes with fluorescence-less artificial intraocular lenses.
AuthorsD Schweitzer, J Haueisen, J L Brauer, M Hammer, M Klemm
JournalBiomedical optics express (Biomed Opt Express) Vol. 11 Issue 10 Pg. 5586-5602 (Oct 01 2020) ISSN: 2156-7085 [Print] United States
PMID33149973 (Publication Type: Journal Article)
Copyright© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement.

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: