Abstract | OBJECTIVE: To evaluate the accuracy and precision of 2 kinds of anti-treponema pallidum (anti-TP) ELISA reagents in our laboratory for detecting the anti-TP in voluntary blood donors, so as to provide the data support for use of ELISA reagents after introduction of chemiluminescene immunoassay (CLIA). METHODS: The route detection of anti-TP was performed by using 2 kinds of ELISA reagents, then 546 responsive positive samples detected by anti-TP ELISA were collected, and the infections status of samples confirmed by treponema pallidum particle agglutination (TPPA) test was identified. The confirmed results of responsive samples detected by 2 kinds of anti-TP ELISA reagents were compared, the accuracy of 2 kinds of anti-TP ELISA reagents was analyzed by drawing ROC and comparing area under curve (AUC), and precision of 2 kinds of anti-TP ELISA reagents was compared by statistical analysis of quality control data from 7.1 2016 to 6.30 2017. RESULTS: There were no statistical difference in confirmed positive rate of responsive samples and weak positive samples between 2 kinds of anti-TP ELISA reagents. The responsive samples detected by 2 kinds of anti-TP ELISA reagents accounted for 85.53%(467/546) of all responsive samples, the positive rate confirmed by TPPA test was 82.87%. 44 responsive samples detected by anti-TP ELISA reagent A and 35 responsive samples detected by anti-TP ELISA reagent B were confirmed to be negative by TPPA test. Comparison of AUC showed that the accuracy of 2 kinds of anti-TP ELISA reagents was more high, the difference between 2 reagents was not statistically significant. The coefficient of variation (CV) of anti-TP ELISA reagent A and B was 14.98% and 18.04% respectively, which met the precision requirement of ELISA test. CONCLUSION: The accuracy and precision of 2 kinds of anti-TP ELISA reagents used in our laboratory are similar, and using any one of anti-TP ELISA reagents all can satisfy the requirements of blood screening.
|
Authors | Nan Gao, Li-Qin Huang, Rui Wang, Jun-Jie Jia, Shuo Wu, Jing Zhang, Hong-Wei Ge |
Journal | Zhongguo shi yan xue ye xue za zhi
(Zhongguo Shi Yan Xue Ye Xue Za Zhi)
Vol. 26
Issue 3
Pg. 905-910
(Jun 2018)
ISSN: 1009-2137 [Print] China |
PMID | 29950241
(Publication Type: Journal Article)
|
Topics |
- Blood Donors
- Enzyme-Linked Immunosorbent Assay
- Humans
- Syphilis Serodiagnosis
- Treponema pallidum
|