Abstract | PURPOSE: The purpose of this study is to explore the feasibility of the use of titanium fiducial markers to minimize the metallic artifact seen with tantalum markers which causes significant distortion on postoperative orbital CT scans. METHOD: We designed and constructed the titanium markers in the shop of Crocker Nuclear Laboratory, UC Davis, CA. The markers were placed on an eyeball phantom. The eyeball was inserted into the Rando phantom in the orbital space. The Rando phantom was imaged with coplanar x rays on Nucletron simulator at UCSF, on digital panel on the eye beam line at CNL eye treatment facility and on CT scanner at UCSF. RESULTS: The titanium markers can be clearly seen on the hard copy of x rays and on digital panel. The CT scan of an orbit using tantalum markers on the right eye and titanium markers on the left eye shows the metal artifact from tantalum markers. Titanium markers show very little distortion on CT images. CONCLUSION: The present study describes these markers and their relative benefit in comparison with tantalum marker, which has been used for localizing ocular tumor for decades.
|
Authors | Inder K Daftari, Jeanne M Quivey, Jennifer S Chang, Kavita K Mishra |
Journal | Medical physics
(Med Phys)
Vol. 45
Issue 3
Pg. 1036-1039
(Mar 2018)
ISSN: 2473-4209 [Electronic] United States |
PMID | 29377168
(Publication Type: Journal Article)
|
Copyright | © 2018 American Association of Physicists in Medicine. |
Chemical References |
|
Topics |
- Choroid Neoplasms
(diagnostic imaging, radiotherapy)
- Feasibility Studies
- Fiducial Markers
- Humans
- Melanoma
(diagnostic imaging, radiotherapy)
- Proton Therapy
- Titanium
- Tomography, X-Ray Computed
|