HOMEPRODUCTSCOMPANYCONTACTFAQResearchDictionaryPharmaSign Up FREE or Login

A sensitive quantitative test strip based point-of-care albuminuria screening assay.

AbstractBACKGROUND:
Chronic kidney disease is a major health problem and the global guidelines require screening of albuminuria. Therefore, affordable and sensitive albuminuria screening tests are needed. We explored the potential of urine strips, generally reported in the ordinal scale, measured on an automatic strip reader for reporting quantitative and sensitive albumin results.
METHODS:
We compared reflectance data of Combur-Test® strips obtained from the Cobas U411 reader (Roche) with albuminuria data from a nephelometer BNII (Siemens) and with protein concentrations from the pyrogallol red method (Modular P, Roche) for 389/328 non-pathologic and pathologic urine samples, respectively.
RESULTS:
Imprecision of the reflectance signal of the Cobas U411 was measured with commercial control material (Bio-Rad). Inter-run coefficients of variations (CVs) for reflectance for levels 1 and 2 were 1.7%/4.9%, respectively, and intra-run CVs were 1.8%/4.2%, respectively. Good agreement was obtained between the albumin concentration of the BNII and the protein strip reflectance data (n=389): Y (10,000/protein reflectance, 1/%)=160+0.132·X (albuminuria BNII, mg/L)-0.0000111·X2 (albuminuria BNII, mg/L); r2=0.921. Lower agreement was found between the protein assay (n=328) and the reflectance (r2=0.831). A calibration curve was made between 11.5 mg/L and 121.5 mg/L. The limit of blank (LOB) was 44.7 mg/L.
CONCLUSIONS:
The present study demonstrates that reflectance data generated by a test strip reader allows for quantitative analysis of albumin. Although the lower limit of the microalbumin range (30 mg/L) cannot be achieved with the dye-binding method, the results are satisfactory for screening purposes.
AuthorsAn-Sofie C Decavele, Tom Fiers, Joris Penders, Joris R Delanghe
JournalClinical chemistry and laboratory medicine (Clin Chem Lab Med) Vol. 50 Issue 4 Pg. 673-8 (Jan 06 2012) ISSN: 1437-4331 [Electronic] Germany
PMID22505559 (Publication Type: Journal Article, Research Support, U.S. Gov't, Non-P.H.S.)
Chemical References
  • Reagent Strips
Topics
  • Adult
  • Albuminuria (diagnosis, urine)
  • Female
  • Humans
  • Hydrogen-Ion Concentration
  • Male
  • Middle Aged
  • Nephelometry and Turbidimetry
  • Point-of-Care Systems (standards)
  • Proteinuria (diagnosis, urine)
  • Reagent Strips (chemistry)
  • Reference Values
  • Young Adult

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: