Abstract |
In order to decrease the morbidity rate after combined radiotherapy of the cervix carcinoma, a tungsten shield 3 and 5 mm thick for the rectum has been developed by the authors which is applied with the ring and pin applicator of the Selectron unit (LDR- and HDR-afterloading). The isodose curves were measured in a plexiglas phantom, and the radiation dose at the reference points was determined by means of a ionization dosimeter. The phantom measurements were performed with the same arrangement of sources as applied in radiotherapy. The measurements showed a dose reduction at point Rmax of 33% (HDR) and 44% (LDR) with the tungsten shield 5 mm thick.
|
Authors | H Hetzel, H Kamleitner, M McCoy, H Frommhold |
Journal | Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]
(Strahlenther Onkol)
Vol. 163
Issue 12
Pg. 782-6
(Dec 1987)
ISSN: 0179-7158 [Print] Germany |
Vernacular Title | Rektumabschirmung für ein Selectron-Ringapplikator-System (HDR- und LDR-Afterloading). |
PMID | 3424077
(Publication Type: English Abstract, Journal Article)
|
Chemical References |
|
Topics |
- Brachytherapy
(instrumentation, methods)
- Carcinoma
(radiotherapy)
- Evaluation Studies as Topic
- Female
- Humans
- Models, Structural
- Radiation Injuries
(prevention & control)
- Radiation Protection
(instrumentation)
- Radiotherapy Dosage
- Rectum
- Tungsten
- Uterine Cervical Neoplasms
(radiotherapy)
|