Abstract | OBJECTIVE: To discuss the influence of setup errors on the accuracy of pelvic cancer in IGRT, analysis setup errors and determine the CTV-to-PTV margins. METHODS: 60 pelvic cancer patients treated with Varian 23IX, all of them were performed by CBCT before and after-correction three times in the first week and after that once a week. Then, to measure the setup errors at X(left-right), Y(superior-inferior), Z(anterior-posterior) axis and E(coronal), F(sagittal), G(axial) rotation directions. RESULTS: 530 scans obtained in all, the setup errors in X, Y, Z, E, F, G were (-0.52 ± 4.18) mm, (0.73 ± 4.86) mm, (-0.36 ± 3.62) mm, (0.14 ± 1.20)degrees, (0.13 ± 1.34)degrees, (0.21 ± 1.73)degrees respectively and were much lower after correction at X, Y, Z axis, besides, CTV-to-PTV margins decrease a lot. CONCLUSION:
|
Authors | Bing Wu, Junxiang He, Jingguo Fu, Changli Ruan |
Journal | Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
(Zhongguo Yi Liao Qi Xie Za Zhi)
Vol. 39
Issue 4
Pg. 299-301
(Jul 2015)
ISSN: 1671-7104 [Print] China |
PMID | 26665956
(Publication Type: Journal Article)
|
Topics |
- Humans
- Pelvic Neoplasms
(radiotherapy)
- Radiotherapy Dosage
- Radiotherapy Planning, Computer-Assisted
- Radiotherapy, Image-Guided
|