An integrated CAD/CAM/robotic milling method for custom cementless femoral prostheses.

Abstract
Aseptic loosening is the primary cause of cementless femoral prosthesis failure and is related to the primary stability of the cementless femoral prosthesis in the femoral cavity. The primary stability affects both the osseointegration and the long-term stability of cementless femoral prostheses. A custom cementless femoral prosthesis can improve the fit and fill of the prosthesis in the femoral cavity and decrease the micromotion of the proximal prosthesis such that the primary stability of the custom prosthesis can be improved, and osseointegration of the proximal prosthesis is achieved. These results will help to achieve long-term stability in total hip arthroplasty (THA). In this paper, we introduce an integrated CAD/CAM/robotic method of milling custom cementless femoral prostheses. The 3D reconstruction model uses femoral CT images and 3D design software to design a CAD model of the custom prosthesis. After the transformation matrices between two units of the robotic system are calibrated, consistency between the CAM software and the robotic system can be achieved, and errors in the robotic milling can be limited. According to the CAD model of the custom prosthesis, the positions of the robotic tool points are produced by the CAM software of the CNC machine. The normal vector of the three adjacent robotic tool point positions determines the pose of the robotic tool point. In conclusion, the fit rate of custom pig femur stems in the femoral cavities was 90.84%. After custom femoral prostheses were inserted into the femoral cavities, the maximum gaps between the prostheses and the cavities measured less than 1 mm at the diaphysis and 1.3 mm at the metaphysis.
AuthorsXi Wen-ming, Wang Ai-min, Wu Qi, Liu Chang-hua, Zhu Jian-fei, Xia Fang-fang
JournalMedical engineering & physics (Med Eng Phys) Vol. 37 Issue 9 Pg. 911-5 (Sep 2015) ISSN: 1873-4030 [Electronic] England
PMID26210779 (Publication Type: Journal Article, Research Support, Non-U.S. Gov't)
CopyrightCopyright © 2015 IPEM. Published by Elsevier Ltd. All rights reserved.
Chemical References
  • Drosophila Proteins
  • sog protein, Drosophila
Topics
  • Animals
  • Arthroplasty, Replacement, Hip
  • Bone-Implant Interface
  • Calibration
  • Computer-Aided Design
  • Drosophila Proteins
  • Femur (diagnostic imaging, surgery)
  • Hip Prosthesis
  • Imaging, Three-Dimensional
  • Models, Biological
  • Precision Medicine (instrumentation, methods)
  • Prosthesis Design (instrumentation, methods)
  • Robotics (methods)
  • Swine
  • Tomography, X-Ray Computed

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: