First experiences with the GORE(®) Septal Occluder in children and adults with atrial septal defects.

AbstractOBJECTIVES:
In this study, we report the first experiences with the GORE(®) Septal Occluder in transcatheter treatment of patients with atrial septal defects (ASD).
METHODS:
Twenty-two patients with secundum ASD were selected for GORE(®) Septal Occluder device closure. Mean age was 25.8 ± 4.6 years (range 4-60), 10 were children below the age of 15. Closure was guided by transeosophageal or intracardiac echocardiography and transthoracic echocardiography was used at follow-up (median time 3 months). All defects were balloon sized and device size was chosen according to an occluder/defect size ratio ≥2.
RESULTS:
The mean ASD size was 11.4 ± 0.5 mm (range 6-15). All GORE(®) Septal Occluder sizes (15, 20, 25 and 30 mm) were used. Six patients had additional defects. The aortic rim was deficient in twelve patients and four patients had a septal aneurysm. Symptoms were seen in 17 patients and 19 had dilated right ventricles. Implantation was successful in all patients with no complications. Delivery was easy and intuitive by the new delivery system. None of the patients had residual shunts at follow-up. Mean procedure time was 38 ± 5 min (range 20-120). All patients had resolution or improvement in symptoms. There were no new onset arrhythmias or AV valve regurgitation and right ventricular dimensions were normalized.
CONCLUSIONS:
The GORE(®) Septal Occluder is an efficient device for closing ASDs up to 15 mm. It is feasible for defects with different anatomy including a deficient aortic rim, septal aneurysm and multiple fenestrations and due to its compliance and conformability it has a presumed low risk of erosion.
AuthorsCamilla Nyboe, Vibeke E Hjortdal, Jens Erik Nielsen-Kudsk
JournalCatheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (Catheter Cardiovasc Interv) Vol. 82 Issue 6 Pg. 929-34 (Nov 15 2013) ISSN: 1522-726X [Electronic] United States
PMID23404677 (Publication Type: Journal Article)
CopyrightCopyright © 2013 Wiley Periodicals, Inc.
Topics
  • Adolescent
  • Adult
  • Cardiac Catheterization (adverse effects, instrumentation)
  • Child
  • Child, Preschool
  • Echocardiography, Doppler, Color
  • Echocardiography, Transesophageal
  • Female
  • Heart Septal Defects, Atrial (diagnostic imaging, physiopathology, therapy)
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Septal Occluder Device
  • Time Factors
  • Treatment Outcome
  • Young Adult

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: