Imaging of acute pulmonary embolism using a dual energy CT system with rapid kVp switching: initial results.

AbstractPURPOSE:
Computed tomography pulmonary angiography (CTPA) is considered as clinical gold standard for diagnosing pulmonary embolism (PE). Whereas conventional CTPA only offers anatomic information, dual energy CT (DECT) provides functional information on blood volume as surrogate of perfusion by assessing the pulmonary iodine distribution. The purpose of this study was to evaluate the feasibility of lung perfusion imaging using a single-tube DECT scanner with rapid kVp switching.
MATERIALS AND METHODS:
Fourteen patients with suspicion of acute PE underwent DECT. Two experienced radiologists assessed the CTPA images and lung perfusion maps regarding the presence of PE. The image quality was rated using a semi-quantitative 5-point scale: 1 (=excellent) to 5 (=non-diagnostic). Iodine concentrations were quantified by a ROI analysis.
RESULTS:
Seventy perfusion defects were identified in 266 lung segments: 13 (19%) were rated as consistent with PE. Five patients had signs of PE at CTPA. All patients with occlusive clots were correctly identified by DECT perfusion maps. On a per patient basis the sensitivity and specificity were 80.0% and 88.9%, respectively, while on a per segment basis it was 40.0% and 97.6%, respectively. None of the patients with a homogeneous perfusion map had an abnormal CTPA. The overall image quality of the perfusion maps was rated with a mean score of 2.6 ± 0.6. There was a significant ventrodorsal gradient of the median iodine concentrations (1.1mg/cm(3) vs. 1.7 mg/cm(3)).
CONCLUSION:
Lung perfusion imaging on a DE CT-system with fast kVp-switching is feasible. DECT might be a helpful adjunct to assess the clinical severity of PE.
AuthorsLucas L Geyer, Michael Scherr, Markus Körner, Stefan Wirth, Paul Deak, Maximilian F Reiser, Ulrich Linsenmaier
JournalEuropean journal of radiology (Eur J Radiol) Vol. 81 Issue 12 Pg. 3711-8 (Dec 2012) ISSN: 1872-7727 [Electronic] Ireland
PMID21420812 (Publication Type: Journal Article)
CopyrightCopyright © 2011 Elsevier Ireland Ltd. All rights reserved.
Topics
  • Acute Disease
  • Adult
  • Aged
  • Algorithms
  • Angiography (methods)
  • Female
  • Humans
  • Image Enhancement (methods)
  • Male
  • Middle Aged
  • Observer Variation
  • Pulmonary Embolism (diagnostic imaging)
  • Radiography, Dual-Energy Scanned Projection (methods)
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed (methods)