HOMEPRODUCTSCOMPANYCONTACTFAQResearchDictionaryPharmaSign Up FREE or Login

Increase in internal defibrillation threshold during acute myocardial infarction.

Abstract
A 67-year-old man suffered an acute anteroseptal myocardial infarction complicated by multiple episodes of ventricular fibrillation, which were not systematically defibrillated by maximum, internal 35-J shocks delivered by an implanted cardioverter defibrillator (ICD). He had suffered from acute inferior myocardial infarction 6 years earlier, complicated with sustained polymorphic ventricular tachycardia (VT). Due to inducibility of sustained VT on an electrophysiologic study, an ICD was implanted. Defibrillation testing performed after healing of anteroseptal infarction was successful with a 10-J safety margin, suggesting that acute myocardial ischemia transiently elevated the internal defibrillation threshold.
AuthorsMasaki Ota, Yoshiaki Kaneko, Tadashi Nakajima, Akihiro Saito, Tadanobu Irie, Toshimitsu Kato, Takafumi Iijima, Toshio Ito, Mamoru Manita, Masahiko Kurabayashi
JournalInternal medicine (Tokyo, Japan) (Intern Med) Vol. 49 Issue 18 Pg. 1975-8 ( 2010) ISSN: 1349-7235 [Electronic] Japan
PMID20847501 (Publication Type: Case Reports, Journal Article)
Topics
  • Aged
  • Defibrillators, Implantable
  • Electrocardiography
  • Humans
  • Male
  • Myocardial Infarction (diagnosis, physiopathology, therapy)

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: