Intermethod variability in TSH-receptor antibody measurement: implication for the diagnosis of Graves disease and for the follow-up of Graves ophthalmopathy.

AbstractBACKGROUND:
We compared the analytical and clinical performance of 3 porcine thyroid receptor antibody (TRAb) methods (1 second- and 2 new third-generation systems) with the conventional TRAb assay based on the human recombinant TSH receptor (hTRAK).
PATIENTS AND METHODS:
We obtained sera from 86 patients with untreated Graves disease (GD) and 71 healthy controls. We measured TRAb concentrations by radioreceptor assay using the hTRAK (Brahms) or the porcine TSH receptor (pRRA) from Beckman-Coulter, by electrochemiluminescence immunoassay (ECLIA) with the Elecsys/Cobas (Roche), and by ELISA using the Medizym TRAb clone (Medipan).
RESULTS:
Between-run assay imprecision was < or =10% and < or =7.6% for hTRAK and ECLIA, but reached 14% and 14.9% for ELISA and pRRA, respectively. Maximal specificity and sensitivity close to 100% were obtained for hTRAK, ECLIA, and ELISA. pRRA failed to detect positive TRAbs in 5 GD patients. Although calibrated against the same reference standard 90/672, the assays displayed a high intermethod variability. The results were significantly higher by ECLIA and lower by ELISA and pRRA compared with hTRAK. Patients with ophthalmopathy had higher TRAb results by ELISA and pRRA than those without eye disease.
CONCLUSIONS:
Second- and third-generation TRAb assays had similar diagnostic sensitivities in the diagnostic evaluation of GD. Despite the use of the same reference standard for calibration, high intermethod variability in TRAb assay results was seen in untreated GD patients. Assay harmonization is necessary for correct interpretation in the follow-up of Graves ophthalmopathy.
AuthorsCatherine Massart, Rémy Sapin, Jacqueline Gibassier, Arnaud Agin, Michèle d'Herbomez
JournalClinical chemistry (Clin Chem) Vol. 55 Issue 1 Pg. 183-6 (Jan 2009) ISSN: 1530-8561 [Electronic] England
PMID19028818 (Publication Type: Evaluation Study, Journal Article)
Chemical References
  • Autoantibodies
  • Receptors, Thyrotropin
  • Recombinant Proteins
Topics
  • Adolescent
  • Adult
  • Aged
  • Animals
  • Autoantibodies (blood, immunology)
  • CHO Cells
  • Cricetinae
  • Cricetulus
  • Female
  • Follow-Up Studies
  • Graves Disease (blood, diagnosis, immunology)
  • Humans
  • Immunoassay
  • Male
  • Middle Aged
  • Radioligand Assay
  • Receptors, Thyrotropin (immunology)
  • Recombinant Proteins (blood, immunology)
  • Reproducibility of Results
  • Sensitivity and Specificity