Smooth nylon foil (SupraFOIL) orbital implants in orbital fractures: a case series of 181 patients.

AbstractPURPOSE:
The SupraFOIL smooth nylon foil implant is a polyamide sheet available in varying thicknesses that has been used in orbital reconstruction following trauma. The authors report their experience with smooth nylon foil implants in the repair of orbital fractures.
METHODS:
A retrospective chart review of patients having undergone repair of orbital fractures between January 1, 1995 and December 31, 2003 was conducted. The Wilcoxon test was used to compare cases with complications and those with no complications.
RESULTS:
Of 282 orbits that underwent fracture repair with alloplastic material, 87 were excluded because non-nylon foil implants were used alone or in combination with nylon foil to repair the orbital fracture. An additional 14 fractures were excluded because of insufficient clinical data or lack of follow-up, for a total of 181 orbits that underwent orbital fracture repair with only nylon foil. A transconjunctival approach was used in 98% cases, and all 181 implants were fixated with at least 1 titanium screw. Average follow-up was 362.8 days. The overall complication rate was 1.7%. One patient had an acute postoperative orbital hemorrhage (0.6% acute complication rate). Two patients had late orbital infections, 683 days and 984 days following repair (1.1% late complication rate).
CONCLUSIONS:
The authors found the smooth nylon foil implant to be safe and effective in orbital fracture repair. The lower rate of complications the authors observed compared with previously reported series may be related to implant fixation. A titanium screw secured just posterior to the orbital rim may decrease long-term complications by providing more stability than implants left without fixation.
AuthorsD J John Park, Daniel C Garibaldi, Nicholas T Iliff, Michael P Grant, Shannath L Merbs
JournalOphthalmic plastic and reconstructive surgery (Ophthalmic Plast Reconstr Surg) 2008 Jul-Aug Vol. 24 Issue 4 Pg. 266-70 ISSN: 0740-9303 [Print] United States
PMID18645428 (Publication Type: Case Reports, Journal Article)
Chemical References
  • Nylons
  • polyamide mesh
Topics
  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Child
  • Child, Preschool
  • Female
  • Fracture Fixation (instrumentation, methods)
  • Humans
  • Incidence
  • Intraoperative Complications
  • Male
  • Middle Aged
  • Nylons
  • Ophthalmologic Surgical Procedures
  • Orbital Fractures (diagnostic imaging, surgery)
  • Orbital Implants
  • Postoperative Complications
  • Plastic Surgery Procedures
  • Retrospective Studies
  • Tomography, X-Ray Computed
  • Treatment Outcome