Bacterial adherence to tantalum versus commonly used orthopedic metallic implant materials.

AbstractOBJECTIVES:
Evaluation of bacterial adhesion to pure tantalum and tantalum-coated stainless steel versus commercially pure titanium, titanium alloy (Ti-6Al-4V), and grit-blasted and polished stainless steel.
DESIGN:
Experimental in vitro cell culture study using Staphylococcus aureus and Staphylococcus epidermidis to evaluate qualitatively and quantitatively bacterial adherence to metallic implants.
METHODS:
A bacterial adhesion assay was performed by culturing S. aureus (ATCC 6538) and S. epidermidis (clinical isolate) for one hour with tantalum, tantalum-coated stainless steel, titanium, titanium alloy, grit-blasted and polished stainless steel metallic implant discs. Adhered living and dead bacteria were stained using a 2-color fluorescence assay. Adherence was then quantitatively evaluated by fluorescence microscopy and digital image processing. Qualitative adherence of the bacteria was analyzed with a scanning electron microscope. The quantitative data were related to the implant surface roughness (Pa-value) as measured by confocal laser scanning microscopy.
RESULTS:
Bacterial adherence of S. aureus varied significantly (p = 0.0035) with the type of metallic implant. Pure tantalum presented with significantly (p < 0.05) lower S. aureus adhesion compared to titanium alloy, polished stainless steel, and tantalum-coated stainless steel. Furthermore, pure tantalum had a lower, though not significantly, adhesion than commercially pure titanium and grit-blasted stainless steel. Additionally, there was a significantly higher S. aureus adherence to titanium alloy than to commercially pure titanium (p = 0.014). S. epidermidis adherence was not significantly different among the tested materials. There was no statistically significant correlation between bacterial adherence and surface roughness of the tested implants.
CONCLUSIONS:
Pure tantalum presents with a lower or similar S. aureus and S. epidermidis adhesion when compared with commonly used materials in orthopedic implants.
CLINICAL IMPLICATION:
Because bacterial adhesion is an important predisposing factor in the development of clinical implant infection, tantalum may offer benefits as an adjunct or alternative material compared with current materials commonly used for orthopedic implants.
AuthorsThomas A Schildhauer, Bruce Robie, Gert Muhr, Manfred Köller
JournalJournal of orthopaedic trauma (J Orthop Trauma) Vol. 20 Issue 7 Pg. 476-84 (Jul 2006) ISSN: 0890-5339 [Print] United States
PMID16891939 (Publication Type: Comparative Study, Journal Article, Research Support, Non-U.S. Gov't)
Chemical References
  • Alloys
  • Stainless Steel
  • Tantalum
Topics
  • Alloys
  • Bacterial Adhesion
  • Orthopedic Procedures
  • Prostheses and Implants
  • Stainless Steel
  • Staphylococcus aureus (physiology)
  • Staphylococcus epidermidis (physiology)
  • Surface Properties
  • Tantalum

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research network!


Choose Username:
Email:
Password:
Verify Password: