The Amplatzer duct occluder: experience in 209 patients.

AbstractOBJECTIVES:
The aim of the study was to assess the safety and efficacy of the Amplatzer ductal occluder (ADO) in transcatheter occlusion of patent ductus arteriosus (PDA).
BACKGROUND:
Transcatheter closure of small to moderate sized PDAs is an established procedure. The ADO is a self-expandable device with a number of salutary features, notably its retrievability, ease of delivery via small 5F to 7F catheters and a range of sizes suitable even for the larger PDAs.
METHODS:
Between November 1997 and August 1999, the ADO was successfully implanted in 205 of 209 patients with PDA. The inclusion criteria for this device occlusion method were patients with clinical and echocardiographic features of moderate to large PDA, weighing > or =3.5 kg as well as asymptomatic adolescents and adults with PDA measuring > or =5.0 mm on two-dimensional (2D) echocardiogram. Occlusion was achieved via the antegrade venous approach. Follow-up evaluations were performed with 2D echocardiogram, color-flow mapping and Doppler measurement of the descending aorta and left pulmonary artery velocity at 24 h and 1, 3, 6 and 12 months after implantation.
RESULTS:
Two hundred and five patients had successful PDA occlusion using this device. The patients were between two months and 50 years (median 1.9) and weighed between 3.4 kg and 63.2 (median 8.4). Infants made up 26% of the total patients. The PDA measured from 1.8 to 12.5 mm (mean 4.9) at the narrowest diameter. Forty-four percent of patients achieved immediate complete occlusion. On color Doppler the closure rates at 24 h and 1 month after implant were 66% and 97%, respectively. At 6 and 12 months all except one patient attained complete occlusion. Device embolization occurred in three patients; in two this was spontaneous, and in the other it was due to catheter manipulation during postimplant hemodynamic measurement. Mild aortic narrowing was seen in an infant.
CONCLUSIONS:
Patent ductus arteriosus occlusion using ADO is safe and efficacious. It is particularly useful in symptomatic infants and small children with relatively large PDA. Embolization can be minimized by selection of appropriate sized devices, and caution should be exercised in infants <5 kg.
AuthorsA A Bilkis, M Alwi, S Hasri, A L Haifa, K Geetha, M A Rehman, I Hasanah
JournalJournal of the American College of Cardiology (J Am Coll Cardiol) Vol. 37 Issue 1 Pg. 258-61 (Jan 2001) ISSN: 0735-1097 [Print] United States
PMID11153748 (Publication Type: Comparative Study, Journal Article)
Topics
  • Adolescent
  • Adult
  • Balloon Occlusion (instrumentation)
  • Cardiac Catheterization (instrumentation)
  • Child
  • Child, Preschool
  • Ductus Arteriosus, Patent (diagnostic imaging, therapy)
  • Embolization, Therapeutic (instrumentation)
  • Female
  • Humans
  • Infant
  • Male
  • Middle Aged
  • Radiography
  • Treatment Outcome

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: