Iridium oxide-coated defibrillation electrode: reduced shock polarization and improved defibrillation efficacy.

AbstractBACKGROUND:
Transvenous implantable cardioverter-defibrillator (ICD) leads are designed to deliver electric shocks to the heart for termination of ventricular dysrhythmias. However, the efficiency of different lead materials has not been well studied. This study compares an ICD lead coated with iridium oxide (IROX), a material that reduces shock-induced polarization, with an otherwise identical, uncoated lead.
METHODS AND RESULTS:
The defibrillation threshold (DFT) was determined in 13 swine with both IROX-coated and uncoated ICD leads paired with an uncoated "can" electrode. The leads were exchanged through a Teflon sheath to reproduce the intracardiac position. The delivered energy DFT of the IROX-coated lead was 15.9+/-5.4 J and was significantly lower than the delivered energy DFT of the uncoated lead (19.1+/-5.1 J; P<.006). The initial lead impedance was equivalent in both leads (IROX, 41.7+/-5.8 omega; uncoated, 41.3+/-4.7 omega; P=NS) at DFT. However, the impedance rose by 7.3+/-2.0 omega during the first phase and by 3.7+/-2 omega during the second phase with the uncoated lead, whereas the corresponding impedance change was 1.0+/-0.3 omega during phase 1 and 1.6+/-0.5 omega during phase 2 (P<.01 each phase) when the IROX-coated lead was used.
CONCLUSIONS:
This study shows that an IROX coating of this lead system significantly lowers the DFT energy in the swine model. The blunting of the impedance rise by the IROX coating that is seen is consistent with a reduction in electrode polarization.
AuthorsM J Niebauer, B Wilkoff, Y Yamanouchi, T Mazgalev, K Mowrey, P Tchou
JournalCirculation (Circulation) Vol. 96 Issue 10 Pg. 3732-6 (Nov 18 1997) ISSN: 0009-7322 [Print] United States
PMID9396477 (Publication Type: Journal Article, Research Support, Non-U.S. Gov't)
Chemical References
  • iridium oxide
  • Iridium
Topics
  • Animals
  • Differential Threshold (physiology)
  • Electric Countershock (instrumentation)
  • Electrodes, Implanted
  • Electrophysiology
  • Iridium
  • Surface Properties
  • Swine