Estimation of Goldmann applanation tonometer intraocular pressure (IOP) from scleral Schiotz IOP values in eyes with type-1 keratoprostheses.

AbstractPURPOSE:
To validate estimation of Goldmann applanation tonometer (GAT) intraocular pressure (IOP) from scleral Schiotz IOP measurements using a regression model in normal eyes and eyes with type-1 keratoprostheses.
METHODS:
In this prospective cross-sectional study, cohort-1 had 253 normal anterior segment eyes, and cohort-2 had 100 eyes with type-1 keratoprostheses. Scleral Schiotz IOP measurements were used (in a non-linear model) to predict GAT IOP values for these eyes. Accuracy of predicted GAT IOP values was assessed using actual GAT IOP values for normal eyes, while for type-1 keratoprosthetic eyes, finger tension (FT) IOP assessments by an experienced glaucoma specialist were used. Primary outcome was agreement between FT IOP (assessed by an experienced glaucoma specialist) and predicted GAT IOP-derived clusters.
RESULTS:
The actual values of GAT IOP measurements in normal eyes (n=253; mean age ±SD, 51.35±15.56 years) ranged between 6 mm Hg and 62 mm Hg (mean=22±10.05 mm Hg). Estimated and actual GAT IOP values for normal eyes were very similar (mean difference=0.05 mm Hg with limits of agreement: -5.39 to 5.5 by Bland-Altman plot). Of the 100 eyes with type-1 keratoprostheses, 68 were classified as having digitally normal IOP, 28 as borderline and 4 as high. The agreement between classification by FT assessment and model-predicted GAT IOP values was substantial (Kappa=0.81, 95% CI 0.69 to 0.93). The accuracy of the model in assessing IOP was found to be 91% (95% CI 0.84 to 0.96).
CONCLUSION:
Scleral Schiotz IOP values along with our predictive model can be an alternative objective method to FT IOP in assessing IOP in eyes with type-1 keratoprostheses.
AuthorsSirisha Senthil, Raghava Chary, Mohammed Hasnat Ali, Nikhil S Choudhari, Chandra Sekhar Garudadri
JournalThe British journal of ophthalmology (Br J Ophthalmol) Vol. 106 Issue 4 Pg. 480-484 (04 2022) ISSN: 1468-2079 [Electronic] England
PMID33293270 (Publication Type: Journal Article, Research Support, Non-U.S. Gov't)
Copyright© Author(s) (or their employer(s)) 2022. No commercial re-use. See rights and permissions. Published by BMJ.
Topics
  • Child
  • Cornea
  • Cross-Sectional Studies
  • Glaucoma
  • Humans
  • Intraocular Pressure
  • Prospective Studies
  • Reproducibility of Results
  • Tonometry, Ocular (methods)