Objective: This study sought to evaluate the safety and efficacy of FIREHAWK, a novel abluminal groove-filled biodegradable polymer sirolimus-eluting stent (SES) in patients with moderate-complex coronary lesions (including patients with small vessel disease, long lesion and multi vessel disease), and to validate the ability of the SYNTAX score (SS) to predict clinical outcomes in patients treated with FIREHAWK stent. Methods: TARGETⅡ was a prospective, multicenter, single-arm clinical trial, a total of 730 patients who underwent percutaneous coronary intervention (PCI) of de novo lesions in native coronary arteries in 24 medical centers in China from August 2011 to February 2012 were enrolled in this study. All patients were exclusively treated with the FIREHAWK stent. Clinical data including patients with diabetes, small vessel disease, long lesion and multi vessel disease were analyzed. The primary composite endpoint was the target lesion failure (TLF) of cardiac death, target vessel-related myocardial infarction (TV-MI), or target lesion revascularization (TLR). The secondary composite endpoint was patient-oriented endpoint (PoCE), a composite of all death, all myocardial in farction (MI), or any repeat revascularization; definite/probable stent thrombosis (ST) (including acute, late, and very late thrombosis) . SS was calculated in lesions with stenosis more than 50% with coronary artery diameter greater than 1.5 mm. Patients were grouped by tertiles of SS (≤7, >7 to ≤12, >12). Follow-up was performed up to 5 years. Results: A total of 730 patients were enrolled in the TARGET Ⅱ trial. The average SS was 10.9±6.9. 683 (93.6%) patients completed 5-year clinical follow-up. The 5-year incidence of TLF was 8.5%(58/683). The incidence of TLF components was as follows: cardiac death 2.0%(14/683), TV-MI 4.4%(30/683), TLR 3.4%(23/683). The incidence of PoCE was 16.4%(112/683). The incidence of definite/probable stent thrombosis was 0.7%(5/683).Multivariable Cox regression analysis showed that the diabetes subgroup (HR=1.123, 95%CI 0.623-2.026, P=0.699), the small vessel disease subgroup (HR=0.909, 95%CI 0.526-1.570, P=0.732), the long lesion subgroup (HR=1.561, 95%CI 0.922-2.640, P=0.097), and the multi vessel disease subgroup (HR=1.062, 95%CI 0.611-1.846, P=0.830) did not increase the HR of TLF compared with the counterpart subgroups. Multivariable Cox regression analysis showed that the hazard of TLF was not increased in the middle and high SS groups as compared with the low SS group (HR=1.203,95%CI 0.607-2.385,P=0.597;HR=1.548,95%CI 0.829-2.892,P=0.171). Conclusions: The 5 years follow-up results of TARGET Ⅱ trial shows that the biodegradable polymer of FIREHAWK stents have long-lasting safety and efficacy for patients with moderate-complex coronary lesions. SS is not the predicting factor for the occurrence of TLF in FIREHAWK treated patients with moderate-complex coronary lesions. Trial Registration Clinical Trials.gov, NCT0141264.
AuthorsL J Xu, Z Gao, Y Song, H H Wang, J J Xu, L J Gao, Y Zhang, L Song, X Y Zhao, J Chen, J Q Yuan, S B Qiao, Y J Yang, B Xu, R L Gao, TARGET Ⅱ investigators
JournalZhonghua xin xue guan bing za zhi (Zhonghua Xin Xue Guan Bing Za Zhi) Vol. 46 Issue 7 Pg. 523-528 (Jul 24 2018) ISSN: 0253-3758 [Print] China
PMID30032542 (Publication Type: Clinical Trial, Journal Article, Multicenter Study)
Chemical References
  • Cardiovascular Agents
  • Polymers
  • Sirolimus
Topics
  • Cardiovascular Agents (administration & dosage)
  • China
  • Coronary Artery Disease (therapy)
  • Drug-Eluting Stents
  • Humans
  • Myocardial Infarction
  • Percutaneous Coronary Intervention
  • Polymers
  • Prospective Studies
  • Sirolimus (administration & dosage)
  • Treatment Outcome