Abstract
The effect of intraocular lens shape on off-axis performance of model pseudophakic eyes was investigated. Image quality criteria included an optimization procedure, spot diagrams, and the modulation transfer function (MTF). The optimum lens shape varies with lens fitting position, corneal asphericity, and pupil size. An earlier investigation showed that the optimum lens shape for on-axis imagery is approximately plano-convex, with the more curved surface facing the cornea. In this investigation, optimum lens shapes for a 10 degrees off-axis position ranged from near this shape to near the reversed shape, i.e., the more curved surface facing the retina. Because the on-axis case should carry more weight and because conditions favoring the reverse form include small pupils for which image quality is fairly insensitive to variations in lens shape, it is suggested that good forms for intraocular lenses range from plano-convex (curved surface facing cornea) to equi-convex.
AuthorsD A Atchison
JournalOptometry and vision science : official publication of the American Academy of Optometry (Optom Vis Sci) Vol. 66 Issue 9 Pg. 579-90 (Sep 1989) ISSN: 1040-5488 [Print] United States
PMID2797711 (Publication Type: Journal Article, Research Support, Non-U.S. Gov't)
Topics
  • Equipment Design
  • Humans
  • Lenses, Intraocular
  • Mathematics
  • Models, Biological
  • Optics and Photonics

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: