Catheter obstruction of intrathecal drug administration system -a case report-.

Abstract
Intrathecal drug administration system (ITDAS) can reduce the side effects while increasing the effectiveness of opioids compared to systemic opioid administration. Therefore, the use of ITDAS has increased in the management of cancer pain and chronic intractable pain. Catheter obstruction is a serious complication of ITDAS. Here, we present a case of catheter obstruction by a mass formed at the side hole and in the lumen. A 37-year-old man suffering from failed back surgery syndrome received an ITDAS implantation, and the ITDAS was refilled with morphine every 3 months. When the patient visited the hospital 18 months after ITDAS implantation for a refill, the amount of delivered morphine sulfate was much less than expected. Movement of the pump rotor was examined with fluoroscopy; however, it was normal. CSF aspiration through the catheter access port was impossible. When the intrathecal catheter was removed, we observed that the side hole and lumen of the catheter was plugged.
AuthorsSeok Myeon Rhee, Eun Joo Choi, Pyung Bok Lee, Francis Sahngun Nahm
JournalThe Korean journal of pain (Korean J Pain) Vol. 25 Issue 1 Pg. 47-51 (Jan 2012) ISSN: 2093-0569 [Electronic] Korea (South)
PMID22259717 (Publication Type: Case Reports)

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: