Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair.

AbstractBACKGROUND:
This study aimed to compare the physicomechanical properties of composite prostheses for laparoscopic ventral hernia repair (LVHR) through standard testing and a proposed classification system.
METHODS:
Seven prostheses (four with absorbable barriers and 3 with nonabsorbable barriers) were evaluated. The barrier layer was removed, after which the area of the interstices and the diameter of the filaments were determined. The barrier layer was left intact during thickness, density, suture retention strength, tear resistance, uniaxial tensile, and ball-burst testing. Specimens were oriented parallel or perpendicular to their longest dimension during testing. One-way analysis of variance (ANOVA) with Tukey's posttest or an unpaired, two-tailed t-test was performed to determine whether differences existed due to mesh or orientation, and a p value<0.05 was considered significant.
RESULTS:
Significant differences were observed between mesh types and due to the orientation of the mesh during testing. Of the absorbable barrier meshes, Bard Sepramesh IP Composite demonstrated the greatest suture retention and tear strengths, followed by C-QUR mesh. Of the permanent barrier meshes, DUALMESH demonstrated the greatest suture retention strength in the perpendicular direction, followed by Bard Composix E/X. DUALMESH and Bard Composix E/X demonstrated equivalent suture retention strength in the parallel direction and equivalent tear resistance in both testing directions. All meshes demonstrated tensile strengths greater than the physiologically relevant range of 16-32 N/cm.
CONCLUSIONS:
This study provided a basic understanding of how the structural aspects of each mesh design influence functionality. Differences between composite barrier prostheses commonly used for LVHR were observed due to barrier type, mesh type, and orientation. A set of standard testing techniques and a classification system also were presented to define fully the properties of these materials.
AuthorsCorey R Deeken, Michael S Abdo, Margaret M Frisella, Brent D Matthews
JournalSurgical endoscopy (Surg Endosc) Vol. 25 Issue 5 Pg. 1541-52 (May 2011) ISSN: 1432-2218 [Electronic] Germany
PMID20976479 (Publication Type: Journal Article)
Topics
  • Absorbable Implants
  • Densitometry
  • Hernia, Ventral (surgery)
  • Humans
  • In Vitro Techniques
  • Laparoscopy
  • Materials Testing
  • Surgical Mesh
  • Sutures
  • Tensile Strength

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: