Postmortem study of femoral osteolysis associated with metal-on-metal articulation in total hip replacement: an analysis of nine cases.

AbstractBACKGROUND:
Improved metal-on-metal articulations were reintroduced in total hip replacement to avoid the osteolysis sometimes seen with conventional ultra-high molecular weight polyethylene bearings. Osteolysis and local lymphocytic infiltration have been reported at revision of some metal-on-metal devices. We report similar and additional results in a study of second-generation metal-on-metal hip implants retrieved post mortem.
METHODS:
Components and surrounding tissues were collected post mortem from seven patients with nine total hip replacements (Zweymüller SL stem with an Alloclassic cup) with Metasul metal-on-metal articulations. All available patient information was recorded. Radiographs of the hips were evaluated for osteolysis. Sections of joint capsule as well as of the femoral implant with surrounding bone were reviewed, and energy-dispersive x-ray analysis was used to evaluate the composition of wear products. The amount of wear was measured for each component (nine femoral heads and eight cup inserts), when possible, by a coordinate measurement machine with use of the dimensional method.
RESULTS:
The patients died between three and ten years after arthroplasty, and six of the seven were asymptomatic at the time of death. One patient, with the highest rate of total wear (i.e., wear of femoral head and acetabular cup; 7.6 microm/yr), had increasing hip pain for one year, and histological analysis confirmed the radiographic findings of osteolysis. For two other patients, histological analysis confirmed the radiographic findings of asymptomatic osteolysis. For three patients, histological analysis revealed osteolysis that had escaped conventional radiographic analysis. Joint capsule tissue showed evidence of metallosis in all hips and local lymphocytic infiltration in eight hips. Energy-dispersive x-ray analysis revealed elements attributable to CoCrMo alloy in all hips and traces of corrosion products in three hips.
CONCLUSIONS:
The postmortem findings of osteolysis and/or lymphocytic infiltration associated with eight clinically well-functioning, low wear devices (a total wear rate of <4 microm/yr) suggest there may be frequent, unappreciated femoral bone loss and local immunological response in patients with second-generation metal-on-metal hip implants. Compared with previous postmortem studies, our findings showed the extent of osteolysis was similar to that with metal-on-polyethylene articulations.
AuthorsMonika Huber, Georg Reinisch, Peter Zenz, Karl Zweymüller, Felix Lintner
JournalThe Journal of bone and joint surgery. American volume (J Bone Joint Surg Am) Vol. 92 Issue 8 Pg. 1720-31 (Jul 21 2010) ISSN: 1535-1386 [Electronic] United States
PMID20660235 (Publication Type: Journal Article)
Chemical References
  • Biocompatible Materials
  • Metals
Topics
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip (adverse effects, instrumentation)
  • Autopsy
  • Biocompatible Materials
  • Equipment Failure Analysis
  • Female
  • Femur
  • Hip Prosthesis (adverse effects)
  • Humans
  • Male
  • Metals
  • Middle Aged
  • Osteolysis (etiology)
  • Prosthesis Failure

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: