Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose.

AbstractUNLABELLED:
The quality of (18)F-FDG PET/CT images of overweight patients is often degraded. We evaluated the effect of optimizing injected dose or acquisition time on the quality of images of overweight patients using lutetium oxyorthosilicate PET/CT with high-performance detector electronics.
METHODS:
We initially retrospectively measured radioactivity concentrations and signal-to-noise ratios (SNRs) in the liver relative to body weight for 80 patients who had undergone (18)F-FDG PET/CT according to our standard protocol (injected dose, 3.7 MBq/kg; acquisition time, 2 min/bed position). The patients were grouped (n = 20 per group) according to baseline body weight as G1 (<or=59 kg), G2 (60-69 kg), G3 (70-84 kg), and G4 (>or=85 kg). We compared the SNRs of G1 with those of G2, G3, and G4 and calculated the ratio squared as a factor to correct the acquisition parameters for overweight patients. We then prospectively enrolled 120 patients according to the same body weight criteria. We multiplied the correction factors to optimize injected doses or acquisition times and defined dose-adjusted groups (n = 20 per group) and time-adjusted groups (n = 20 per group). G2 dose was defined as 5.59 +/- 0.19 MBq/kg, G3 dose as 7.29 +/- 0.33 MBq/kg, and G4 dose as 8.88 +/- 0.43 MBq/kg. G2 time was defined as 3 min/bed position, G3 time as 4 min/bed position, and G4 time as 5 min/bed position.
RESULTS:
Although liver activities did not significantly differ among G1 through G4 irrespective of patient weight, SNR progressively decreased as patient weight increased. The liver activities of G2 dose, G3 dose, and G4 dose were, respectively, 1.4-, 1.9-, and 2.5-fold higher than those of the baseline counterparts. Nevertheless, the increased liver activities of G2 dose, G3 dose, and G4 dose did not significantly affect SNR, compared with the baseline groups. In contrast, the SNR of G4 time was significantly higher than that of G4.
CONCLUSION:
Our findings suggest that the quality of images acquired from heavier patients can be maintained only by scanning for longer periods. Increasing the dose per kilogram of body weight did not improve the quality of lutetium oxyorthosilicate PET/CT images.
AuthorsYoko Masuda, Chisato Kondo, Yuka Matsuo, Masataka Uetani, Kiyoko Kusakabe
JournalJournal of nuclear medicine : official publication, Society of Nuclear Medicine (J Nucl Med) Vol. 50 Issue 6 Pg. 844-8 (Jun 2009) ISSN: 0161-5505 [Print] United States
PMID19443586 (Publication Type: Comparative Study, Journal Article)
Chemical References
  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18
Topics
  • Female
  • Fluorodeoxyglucose F18
  • Humans
  • Liver (diagnostic imaging)
  • Male
  • Overweight (diagnostic imaging)
  • Positron-Emission Tomography (methods)
  • Radiation Dosage
  • Radiopharmaceuticals
  • Tomography, X-Ray Computed (methods)