Abstract | STUDY DESIGN: OBJECTIVES: To clarify technical requirements and test the hypothesis that placing two linear arrays in parallel, thereby doubling the number of contacts, improves outcome. SUMMARY OF BACKGROUND DATA: METHODS: Acting as their own controls, 20 patients who passed screening with single, 4-contact electrodes received permanent dual, 4-contact electrodes with 7- or 10-mm intercontact distances at the same vertebral level(s). We quantified and compared the technical and clinical results of the single and dual electrodes, adjusting stimulation parameters to specific psychophysical thresholds. RESULTS: Single electrodes provided significant (P < 0.01) advantages in patient- and computer-calculated ratings of pain coverage by paresthesias and in the scaled amplitude necessary to cover the low back, compared with dual 7-mm electrodes. Slight advantages without statistical significance were observed for the single over the dual 10-mm electrodes. Amplitude requirements were significantly lower for the single electrode than for either dual electrode. At long-term follow-up, 53% of patients met the criteria for clinical success. CONCLUSIONS: While we observed disadvantages for dual electrodes in treating axial low back pain, we achieved technical success with single or dual electrodes in most patients and maintained this success clinically with dual electrodes in 53%.
|
Authors | Richard B North, David H Kidd, John Olin, Jeffrey M Sieracki, Farrokh Farrokhi, Loredana Petrucci, Protagoras N Cutchis |
Journal | Spine
(Spine (Phila Pa 1976))
Vol. 30
Issue 12
Pg. 1412-8
(Jun 15 2005)
ISSN: 1528-1159 [Electronic] United States |
PMID | 15959371
(Publication Type: Comparative Study, Controlled Clinical Trial, Journal Article, Research Support, Non-U.S. Gov't)
|
Topics |
- Adult
- Aged
- Electrodes, Implanted
- Equipment Design
- Female
- Humans
- Low Back Pain
(physiopathology, therapy)
- Male
- Middle Aged
- Pain, Intractable
(physiopathology, therapy)
- Prospective Studies
- Spinal Cord
(physiopathology)
- Transcutaneous Electric Nerve Stimulation
(instrumentation, methods)
- Treatment Outcome
|