Intravascular hemolysis in patients with new-generation prosthetic heart valves: a prospective study.

AbstractOBJECTIVE:
A prospective clinical study was designed to assess the frequency and severity of intravascular hemolysis in patients with new-generation, normally functioning prosthetic heart valves.
METHODS:
Hemolysis was evaluated in 172 patients with a mechanical prosthesis (53 CarboMedics and 119 Sorin Bicarbon) and in 106 patients with a bioprosthesis (15 St Jude Medical Toronto, 19 Baxter Perimount, and 72 Medtronic Mosaic) in the aortic position, mitral position, or both. Aortic valve replacement was performed in 206 patients, mitral valve replacement in 59 patients, and double valve replacement in 13 patients. The presence of hemolysis was assessed on the basis of the level of serum lactic dehydrogenase and serum haptoglobin and the presence and amount of reticulocytes and schistocytes in the peripheral blood. Severity of intravascular hemolysis was estimated on the basis of serum lactic dehydrogenase. Clinical, echocardiographic, and hematologic evaluations were performed 1, 6, and 12 months after discharge.
RESULTS:
None of the 278 patients experienced decompensated anemia, whereas at 12 months, mild subclinical hemolysis was identified in 49 patients, 44 (26%) with a mechanical prosthesis and 5 (5%) with a bioprosthesis (P <.001). At multivariate analysis, independent predictors of the presence of subclinical hemolysis were mitral valve replacement (P <.001), use of a mechanical prosthesis (P =.002), and double valve replacement (P =.02). Frequency of hemolysis in patients with stented aortic bioprostheses was 3%, whereas it was absent in those with stentless valves. Among mechanical valve recipients, double versus single valve replacement (P =.04) and mitral versus aortic valve replacement (P =.05) were correlated with the presence of hemolysis; double valve recipients also showed a more severe degree of hemolysis (P =.03). In patients with a Sorin Bicarbon prosthesis, hemolysis was less frequent (22% vs 34%, P =.09) and severe (P <.001) than in those with a CarboMedics prosthesis.
CONCLUSIONS:
In normally functioning prosthetic heart valves, subclinical hemolysis is a frequent finding. A low incidence of hemolysis is found in stented biologic prostheses, and it is absent in stentless aortic valves. Modifications of valve design may contribute to minimize the occurrence of hemolysis in mechanical prostheses.
AuthorsGianclaudio Mecozzi, Aldo D Milano, Marco De Carlo, Flavia Sorrentino, Stefano Pratali, Carmela Nardi, Uberto Bortolotti
JournalThe Journal of thoracic and cardiovascular surgery (J Thorac Cardiovasc Surg) Vol. 123 Issue 3 Pg. 550-6 (Mar 2002) ISSN: 0022-5223 [Print] United States
PMID11882830 (Publication Type: Journal Article)
Chemical References
  • Haptoglobins
  • L-Lactate Dehydrogenase
Topics
  • Adult
  • Aged
  • Aged, 80 and over
  • Aortic Valve
  • Bioprosthesis
  • Female
  • Haptoglobins (analysis)
  • Heart Valve Prosthesis (adverse effects)
  • Hemolysis
  • Humans
  • L-Lactate Dehydrogenase (blood)
  • Male
  • Middle Aged
  • Mitral Valve
  • Prospective Studies
  • Prosthesis Design

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: