Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes.

AbstractPURPOSE:
To compare corneal thickness measurements made by ultrasonic and slit-scanning techniques in normal eyes and in eyes after laser in situ keratomileusis (LASIK).
SETTING:
Corneal Diseases and Excimer Laser Research Unit, University of Dundee, Dundee, Scotland.
METHODS:
Central corneal thickness (CCT) was measured in 101 eyes of 59 normal subjects and in 30 eyes of 21 post-LASIK patients. Measurements were made with an Orbscan slit-scanning elevation topographer and immediately afterward with an ultrasound pachymeter.
RESULTS:
The difference in mean CCT between ultrasound (538.0 microm +/- 36.7 [SD]) and Orbscan (566.6 +/- 40.7 microm) pachymetry was statistically significant (P <.001) in the normal eyes; the Orbscan measurement was approximately 28 microm higher than that of the ultrasound pachymeter. The difference in mean CCT between the ultrasound and the slit-scanning techniques was also statistically significant in the post-LASIK eyes (mean values 475.3 +/- 50.3 microm and 461.9 +/- 74.2 microm, respectively; P <.0001). Differences in CCT in individual subjects were much more variable in the post-LASIK eyes than in the normal eyes. The Bland and Altman method for assessing clinical agreement between 2 instruments showed that in 95% of cases, the CCT measurements with both instruments would be within 65 microm in normal eyes and 150 microm in post-LASIK eyes.
CONCLUSION:
Central corneal thickness measurements were, on average, 28 microm higher with the Orbscan than with the ultrasound pachymeter in normal eyes and 13 microm lower in post-LASIK eyes. The degree of variability within each group indicated that these 2 techniques are not clinically comparable, precluding interchangeable use of their data in planning or assessing corneal surgery.
AuthorsH S Chakrabarti, J P Craig, A Brahma, T Y Malik, C N McGhee
JournalJournal of cataract and refractive surgery (J Cataract Refract Surg) Vol. 27 Issue 11 Pg. 1823-8 (Nov 2001) ISSN: 0886-3350 [Print] United States
PMID11709257 (Publication Type: Comparative Study, Journal Article, Research Support, Non-U.S. Gov't)
Topics
  • Adult
  • Cornea (diagnostic imaging, surgery)
  • Corneal Topography
  • Humans
  • Keratomileusis, Laser In Situ
  • Myopia (diagnostic imaging, surgery)
  • Prospective Studies
  • Surveys and Questionnaires
  • Ultrasonography