Abstract |
L-775,606 is under investigation as a selective 5-hydroxytryptamine 1D agonist for the treatment of migraine. A reliable and sensitive method for the analysis of L-775,606 in plasma was required in order to support preclinical evaluation of this compound. A semi-automated sample preparation method using the Beckman Biomek 2000 workstation to perform all liquid handling tasks has been established. The sample analysis was performed using HPLC-MS-MS with a cycle time of 3.5 min per sample. Intra- and inter-day assay accuracy and precision are excellent with a calibration range of 1-2000 ng/ml and a reproducible limit of quantification of 1 ng/ml.
|
Authors | K L Locker, D Morrison, A P Watt |
Journal | Journal of chromatography. B, Biomedical sciences and applications
(J Chromatogr B Biomed Sci Appl)
Vol. 750
Issue 1
Pg. 13-23
(Jan 05 2001)
ISSN: 1387-2273 [Print] Netherlands |
PMID | 11204213
(Publication Type: Journal Article)
|
Chemical References |
- 1-(3-(5-(1,2,4-triazol-4-yl)-1H-indol-3-yl)propyl)-4-(2-(3-fluorophenyl)ethyl)piperazine
- Indoles
- Piperazines
- Receptor, Serotonin, 5-HT1D
- Receptors, Serotonin
- Serotonin Receptor Agonists
|
Topics |
- Animals
- Automation
- Calibration
- Chromatography, Liquid
(methods)
- Indoles
(blood)
- Male
- Mass Spectrometry
(methods)
- Piperazines
(blood)
- Rats
- Rats, Sprague-Dawley
- Receptor, Serotonin, 5-HT1D
- Receptors, Serotonin
(drug effects)
- Reproducibility of Results
- Sensitivity and Specificity
- Serotonin Receptor Agonists
(blood)
|