HOMEPRODUCTSCOMPANYCONTACTFAQResearchDictionaryPharmaSign Up FREE or Login

[Hemolysis, a rare complication after percutaneous closure of an atrial septal defect].

Abstract
Haemolysis after closure of an atrial septal defect with an Amplatzer device is a rare complication. The suggested mechanism is the association of a residual shunt with flow of tricuspid regurgitation, impinging on the inferior edge of the prosthesis, factors which should be taken into consideration during the implantation of this type of device.
AuthorsV Lambert, E Belli, J D Piot, C Planché, J Losay
JournalArchives des maladies du coeur et des vaisseaux (Arch Mal Coeur Vaiss) Vol. 93 Issue 5 Pg. 623-5 (May 2000) ISSN: 0003-9683 [Print] France
Vernacular TitleL'hémolyse, une complication exceptionnelle après fermeture percutanée de communication interauriculaire.
PMID10858861 (Publication Type: Case Reports, English Abstract, Journal Article)
Topics
  • Adolescent
  • Echocardiography
  • Embolization, Therapeutic
  • Female
  • Heart Septal Defects, Atrial (therapy)
  • Hemolysis
  • Humans
  • Prosthesis Failure
  • Stents (adverse effects)